DFM 5.1 Doppler Flow Meter
SIMULATION
Exercises the 4-20mA (0-5V) output, digital display and control relays.
Test Select Maximum and press to simulate maximum Flow or Velocity and
to output 20mA (5V) to the analog channel.
Select Minimum and press to simulate minimum Flow or Velocity and
to output 4mA (0V) to the analog channel.
To simulate measurements between minimum and maximum set Test to
Actual and then enter for the flow measurement. The analog output and
control relays will respond to the simulated value.