• The unit should be cleaned with a soft, dry cloth.
• Use a soft, moistened cloth and soap to clean the arm cuff.
• Substitution of a cuff different from that supplied might result in measurement error.
• We will provide circuit diagrams, component part lists, descriptions, calibration instructions, or
other information to assist service personnel to repair parts.
• The service personnel need to be the personnel who is trained by the manufacture and has got
the relevant qualification certificate.
• The accuracy of this blood pressure monitor has been carefully tested and is designed for a long
service life. We recommend that the performance be checked every 2 years and after
maintenance and repair, by utilizing the manometer mode and verifying the accuracy of the
manometer at least at 50 mmHg (6,7kPa) and 200 mmHg(26,7kPa). Please contact the local
distributor for details.
Specifications
Statement
• The unit is intended to be used to measure blood pressure (systolic and diastolic) and heart rate
from the upper arm by using the oscillometric method.
• The unit is intended for using in only adult population, not applied to the other populations such as
neonatal baby.
• It can’t be used while the wrist (arm) has bleeding or wound to avoid the blood flowing from the
wound in pressurizing.
• The device, part and batteries have to be disposed of waste correctly at the end of the usage.
• Please follow Local Ordinances or Regulations for disposal.
• Applied part: CUFF.
• Protection Class: Internally powered equipment.
• Applied Part Type: Type BF.
• Moisture Protection: IPX0, continue operation.
altitude<2000m;
0 ~ 299 mmHg (0~39.9 kPa)
Accuracy
LCD Indication
3 Digits Display of mmHg or kPa
Memory/Heartbeat/Low Battery/IHB, etc.
Automatic by Internal Pump
Deflation Automatic by Air Valve
Memory Function 2 x 90 Sets of Measurement Values
3 AA Alkaline Batteries; D.C.6V
Automatic Power-Off In 2 Minutes
Approx. 290g (Batteries not included)
Approx 105 g (Batteries not included)
Operating Environment
Humidity 15 ~ 90% RH(Noncondensing)
Transport Storage Environment
Temperature -20~60 °C (-4 ~ 140 °F)
10 ~ 95% RH (Noncondensing)
Note: Subject to technical modification without prior notice.
871125208143.indd 16 14-03-18 14:53