Guildline Autosal 8400B - Page 64

Default Icon
71 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 12
TM8400B-L-00
1 November 2004
12-3
PROBLEM
SYMPTOM
PROBABLE CAUSE
3.
Unstable
Reading
Conductivity reading
fluctuates excessively
Salinity Gradients in sample
- ensure sample is well mixed.
Sample is too hot or cold
- sample temperature must be stabilized to
bath temperature in heat exchanger
- sample temperature should not exceed 5°C
of bath temperature at full flow rate.
Sample flow rate too high.
Temperature Drift.
- check temperature control circuit, and bath
temperature.
Cell contamination - bubbles in cell arms
- check cell visually for contamination or
improper arm filling, remove if necessary
(see operating manual).
Ground path through sample exhaust.
- waste sample hose must not make contact
with external devices such as waste buckets,
water path must be broken.
Noisy Conductivity Circuit.
- check circuit boards for salt contamination
or defective components.
- external Radio Frequency (RF) or
Electromagnetic Interference (EMI) sources
present (with 3V/M source, noise will
typically degrade to ±0.2% fullscale, or
display reading ±0.0500).
- poor grounding of the instrument to the
laboratory grounding system via the line
cord ground pin.
Frequency beat of conductivity circuit
chopper and/or meter clock with line
frequency.
- check frequencies in calibration procedure.
Cell polarizing
- see step 5.