EasyManua.ls Logo

Guildline Autosal 8400B - Page 66

Default Icon
71 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 12
TM8400B-L-00
1 November 2004
12-5
PROBLEM
SYMPTOM
PROBABLE CAUSE
10.
Unstable
Temperature
Heater lamps on
continuously
Bath Temperature set too high (maximum
4°C above ambient).
Sample temperature too low.
Control circuit malfunction
- see step 4.
11.
Bubbles in cell
Bubbles in cell which
do not disappear after
repeated flushings
Foreign matter on cell wall.
- try flushing repeatedly with a wetting agent.
Table 12-1: Troubleshooting Guide