GDS-2000A Series User Manual 
148 
AC coupling. Blocks DC components 
from the trigger circuits. 
High frequency filter above 100kHz  
Low frequency filter below 5kHz 
DC coupling with low sensitivity to 
reject noise. 
Slope 
(Edge, Delay, Rise 
& Fall) 
Trigger on a rising edge. 
Trigger on a falling edge. 
Either. (either rising or falling edge) 
(Edge, Delay, Rise & Fall trigger type 
only) 
Trigger Level 
(Edge, Delay) 
Adjusts the trigger 
manually using the 
Trigger LEVEL knob. 
Sets the trigger level to 1.4V, suitable 
for triggering on TTL signals. 
Sets the trigger to -1.3V. This is 
suitable for ECL circuits. 
Sets the trigger level to 
50% of the waveform 
amplitude.  
Set the holdoff time to the minimum. 
Sets the delay time (10ns ~ 10s) 
between the trigger event and the real 