EasyManua.ls Logo

GW Instek GDS-3352A - Using Advanced Delay Trigger

GW Instek GDS-3352A
209 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GDS-3000 Series User Manual
134
12. Set the external trigger level using
the side menu.
Range
00.0V~ 5 screen divisions
Set to TTL 1.4V
Set to ECL -1.3V
Set to 50%
Using Advanced Delay Trigger
The EXT trigger source is always used as the delay
triggering source.
1. Press the trigger Menu key.
Menu
2. Press Type from the lower bezel
menu.
3. Select Delay from the side menu.
The delay + edge trigger indicator
appears at the bottom of the
display.
From left: external source, slope, trigger level,
delay + external source
4. To set the delay press Delay from
the bottom bezel.

Table of Contents