6.9 The SHIFT Key and Function Keys 
The                  button is used to enable the secondary function of certain function 
keys with blue printed letters. After pressing the                button, The SHIFT LED 
will light up, only the buttons with blue printed letters are working. To release the 
secondary function by pressing                again.   
 
  The Secondary Functions 
1.
                +    Return to the default status of GFG-3015. 
2.
                +  Retrieve the setup parameters from memory; 
3.
                +    Display sweep center frequency and get into enter mode.
                +    Change the Modulation source. 
5.
                +    Set Frequency modulation function. 
6.
                +    Choose the Modulation type. 
7.
                +    Display Sweep stop frequency and get into enter mode. 
8.
                +    Set LOG sweep function. 
9.
                +    Choose the Counter signal source. 
 
6.10 Setup of LIN or LOG Sweep 
GFG-3015  can  adopt  frequency  to  sweep  its  function  output  for  triangle  and 
ramp waves. The type of sweep can be set as linear or log sweep. 
 Select a main waveform by using                  button. 
 Select a main output amplitude by using                  button. 
 Select the sweep mode by using                  button. 
 Press                to set up sweep RATE value (Range from 0.01Hz to 10kHz).