REMOTE CONTROL
135
Syntax
Query Syntax
MANU:ACW:CLOSet<NR2>
MANU:ACW:CLOSet?
Parameter/
Return parameter
<NR2>
0.000 ~ 041.9 (GPT-98XX)
0.000 ~ 109.9 (GPT-99XX/99XXA)
Example
MANU:ACW:CLOS 20.0
Sets the ACW LO SET current to 20 mA.
MANU:ACW:TTIMe
Set
Query
Description
Sets or returns the ACW test time in seconds. The
test must first be in ACW mode before this
command can be used.
Note: A “TIME ERR” will result if the Ramp Time
+ Test Time is 240 seconds when the HI SET limit
is over 30mA (GPT-98XX) or over 80mA (GPT-
99XX/99XXA). This applies to the ACW function
only.
In special MANU mode, the TIMER can be turned
off.
Syntax
Query Syntax
MANU:ACW:TTIMe {<NR2>|OFF}
MANU:ACW:TTIMe?
Parameter
<NR2>
OFF
0.5 ~ 999.9 seconds
TIMER OFF (special MANU mode).
Return parameter
<NR2>
TIME OFF
0.5 ~ 999.9 seconds
TIMER is OFF (special MANU
mode).
Example
MANU:ACW:TTIM 1
Sets the ACW test time to 1 second.