REMOTE CONTROL
137
MANU:ACW:ARCCurrent
Set
Query
Description
Sets or returns the ACW ARC current value in mA.
ARC must be enabled before the ARC current can
be set. The test must first be in ACW mode before
this command can be used.
ARC current uses the same range as the HI SET
value. The ARC current is limited to 2X the HI SET
value.
Syntax
Query Syntax
MANU:ACW:ARCCurrent <NR2>
MANU:ACW:ARCCurrent?
Parameter/
Return parameter
<NR2>
1.000 ~ 080.0 (GPT-98XX)
2.000 ~ 200.0 (GPT-99XX/99XXA)
Example
MANU:ACW:ARCC 0.04
Sets the ACW ARC value to 0.04 mA.
MANU:DCW:VOLTage
Set
Query
Description
Sets or returns the DCW voltage in kV. The test
must first be in DCW mode before this command
can be used.
Note: A “DC Over 50W” error will result if the
DCW Voltage X HI SET value is > 50 watts (GPT-
98XX).
Note: A “DC Over 100W” error will result if the
DCW Voltage X HI SET value is > 100 watts (GPT-
99XX/99XXA).
Syntax
Query Syntax
MANU:DCW:VOLTage <NR2>
MANU:DCW:VOLTage?
Parameter/
Return parameter
<NR2>
0.050 ~ 6.100 (kV)
Example
MANU:DCW:VOLT 6
Sets the DCW voltage to 6 kV.
w ww . . co m
information@itm.com1.800.561.8187

Other manuals for GW Instek GPT-9902A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-9902A and is the answer not in the manual?

GW Instek GPT-9902A Specifications

General IconGeneral
BrandGW Instek
ModelGPT-9902A
CategoryTest Equipment
LanguageEnglish

Related product manuals