REMOTE CONTROL
141
Example
MANU:DCW:ARCC 10
Sets the DCW ARC value to 10mA.
MANU:IR:VOLTage
Set
Query
Description
Sets or returns the IR voltage in kV. The test must
first be in IR mode before this command can be
used.
Syntax
Query Syntax
MANU:IR:VOLTage <NR2>
MANU:IR:VOLTage?
Parameter/
Return parameter
<NR2>
0.05 ~ 1 (0.05kV to 1kV: steps of .05)
*GPT-99XX/99XXA also includes a
0.125kV point.
Example
MANU:IR:VOLT 1
Sets the IR voltage to 1 kV.
MANU:IR:RHISet
Set
Query
Description
Sets or returns the IR HI SET resistance value in
M(GPT-98XX) or GΩ. The test must first be in IR
mode before this command can be used.
Syntax
Query Syntax
MANU:IR:RHISet <NR1>|NULL
MANU:IR:RHISet?
Parameter/
Return parameter
<NR1>
GPT-98XX only:
2 ~ 9999 (unit = M)
GPT-99XX/GPT-99XXA only:
Format A: 0.002 ~ 50.00 (unit = G)
Format B: 0.002G ~ 50.00G
Format C: 2M ~ 50000M
NULL
Sets the HI SET value to ∞.
Example
(GPT-98XX)
MANU:IR:RHIS 10
Sets the IR HI SET resistance to 10 M.