REMOTE CONTROL
143
MANU:IR:REF
Set
Query
Description
Sets or returns the IR reference value in M(GPT-
98XX) or GΩ. The test must first be in IR mode
before this command can be used.
The reference value must be lower than the HI SET
value.
Syntax
Query Syntax
MANU:IR:REF <NR1>
MANU:IR:REF?
Parameter/
Return parameter
<NR1>
GPT-98XX only:
0000 ~ 9999 (unit = M)
GPT-99XX/GPT-99XXA only:
Format A: 0 ~ 50.00 (unit = G)
Format B: 0G ~ 50.00G
Format C: 0M ~ 50000M
Example
(GPT-98XX)
MANU:IR:REF 900
Sets the IR reference to 900 M.
Example
(GPT-99XX/
99XXA)
MANU:IR:REF 0.900
Sets the IR reference to 900 M.
MANU:GB:CURRent
Set
Query
Description
Sets or returns the GB current in A. The test must
first be in GB mode before this command can be
used.
Syntax
Query Syntax
MANU:GB:CURRent <NR2>
MANU:GB:CURRent?
Parameter/
Return parameter
<NR2>
3.00~33.00 (GPT-98XX)
3.00~33.00 (GPT-99XX)
Example
MANU:GB:CURR 3.00
Sets the GB current to 3.00A.
w ww . . co m
information@itm.com1.800.561.8187

Other manuals for GW Instek GPT-9902A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-9902A and is the answer not in the manual?

GW Instek GPT-9902A Specifications

General IconGeneral
BrandGW Instek
ModelGPT-9902A
CategoryTest Equipment
LanguageEnglish

Related product manuals