To reset setting values (MEAS SETUP) and offset values
(OPEN SHORT) to factory default, use the DEFAULT
SET setting. The device can be quickly restored to its
factory settings.
Default settings:
(MEAS SETUP):
FUNC: Cp-D
FREQ: 1kHz
LEVEL: 1V
RANGE: AUTO
TRIG: INT
SPEED: SLOW
SRC RES: 100Ω
BIAS: OFF
AUTO LCZ: OFF
DELAY: 0ms
ALC: OFF
AVG: 1
MON1: OFF
MON2: OFF
(OPEN SHORT):
OPEN TEST: ON
SHORT TEST: ON
SPOT: OFF
To reset setting values to factory default:
Step 1. Press the [Measure] or [Setup] key.
Step 2. Press the [SYSTEM] bottom soft-key.
Step 3. Use the cursor key to select the
[DEFAULT SET] field.
Step 4. Use the soft-keys to set.