LCR-800 Series User Manual
34
Time Minimum
Value
Maximum
Value
T1 Trigger Pulse Width 5 us
T2 Measurement Start
Delay Time
140 us
T3 /EOM Delay Time
After Data Output
5 us
T4 Calculation and
binning Time
6 ms
SPEED Analog Measurement Time(T5)
Trigger Wait Time
After /EOM
Output (T6)
0.012kHz 0.1kHz 0.12kHz 1kHz 10kHz 100kHz OFF BIN VALUE
SLOW 817ms 901ms 901ms 903ms 873ms 873ms 2ms 4ms 16ms
MEDIUM 817ms 125ms 105ms 59ms 53ms 53ms 2ms 4ms 16ms
FAST 817ms 125ms 103ms 27ms 17ms 17ms 2ms 4ms 16ms
ACCURACY
MODEL
FAST MEDIUM SLOW
LCR-827
LCR-829
LCR-826
0.5% 0.2% 0.1%
* DATA: /O_BIN_1-_BIN_13,/O_P_OVER,/O_S_OVER,/O_S_REJ,/O_P_HI,/O_P_LO