APPENDIX
259
PEL-2040 PEL-2041
CHANNELS 1 1
RANGE Low High Low High
POWER 350W 350W 350W 350W
CURRENT 0~7A 0~70A 0~1A 0~10A
VOLTAGE 1~80V 2.5~500V
MIN.OPERATING
VOLTAGE(dc)
1.0V at 7A 1.0V at 70A
2.5V at 1A 2.5V at 10A
STATIC MODE
CONSTANT CURRENT MODE
Range 0~7A 0~70A 0~1A 0~10A
Resolution 1mA 10mA 0.2mA 2mA
Accuracy
±(0.1%set +
0.1%F.S)
±(0.1%set +
0.2%F.S)
±(0.1%set +
0.1%F.S)
±(0.1%set +
0.2%F.S)
CONSTANT RESISTANCE MODE
Range
0.025~100(350W/16V)
1.25~5K(350W/80V)
1.25~5K(350W/125V)
50~200K(350W/500V)
Resolution 12bits 12bits
Accuracy
100 : ±(0.2%set() + 0.1)
5K: ±(0.1%set() + 0.02)
with 2.5V at input
5K : ±(0.2%set() + 0.02)
200K: ±(0.1%set() +
0.01) with 5V at input
CONSTANT VOLTAGE MODE
Range 1~80V 2.5~500V
Resolution 20mV 100mV
Accuracy ±(0.05%set + 0.1%F.S.) ±(0.05%set + 0.1%F.S.)
Range 0~70A 0~10A
Resolution 10mA 2mA
Accuracy ±(0.1%set + 0.2%F.S.) ±(0.1%set + 0.2%F.S.)
DYNAMIC MODE
CONSTANT CURRENT MODE
T1&T2 0.025mS ~ 10mS / Res :
1uS
10mS ~ 30S / Res : 1mS
0.025mS ~ 10mS / Res :
1uS
10mS ~ 30S / Res : 1mS
Accuracy 1uS / 1mS ± 100ppm 1uS / 1mS ± 100ppm
Slew Rate 0.001 ~
0.28A/uS
0.01 ~
2.8A/uS
0.16 ~
40mA/uS
1.6 ~
400mA/uS
Slew Rate
Resolution
0.001A/uS 0.01A/uS 0.16mA/uS 1.6mA/uS
Slew Rate setting
Accuracy
±(10% +
15us)
±(10% +
15us)
±(10% +
15us)
±(10% +
15us)