EasyManua.ls Logo

GW Instek PEL-2000 Series - Page 62

GW Instek PEL-2000 Series
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PEL-2000 Series User Manual
62
Range
There are two selectable ranges for constant current
mode: high and low range.
Low range has a higher resolution, but a lower
range. If the current exceeds the Low Range, High
range must be used.
Static Functions
Static mode tests the stability of the voltage output
from a power source. Single channel load modules
can have two 2 current levels A (A Vaue) & B (B
Value). A & B have the same range. Pressing the
A/B key on the module load will cycle through the
A and B states. Alternatively, the mainframe can
select A or B Value.
Dual channel load modules only have one current
level (A Value) per channel in static mode.
Static Mode:
Single Channel
Load module.
Dynamic
Functions
Dynamic load functions allow you to set load
levels (Level1, Level2), load time (Timer1, Timer2),
and the slew rate (rising, falling). Depending on
the settings, the load will switch automatically
between levels 1 and 2.
Dynamic loading can be used for charge discharge
cycle testing etc.
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for GW Instek PEL-2000 Series