Specifications
Specifications are subject to change without notice.
Table 1 Probe specifications
8310 / 8315 8311 / 8316
Applications Pure and ultra pure water Moderately conductive solutions
K (cm
-1
) 0.01 0.1
Accuracy < 2% < 2%
Transmitter measurement range 0.01 to 200 μS.cm
-1
0.1 μS to 2 mS.cm
-1
Pt100 temperature response (t
90%)
< 30 seconds < 45 seconds
Maximum temperature (°C) 125 (8310)
150 (8315)
125 (8311)
150 (8316)
Maximum pressure (bars) 10 (8310)
25 (8315)
10 (8311)
25 (8316)
Sample connection ¾ inch NPT ¾ inch NPT
8312 / 8317 8394
Applications Waste water and drinking water Food and pharmaceutical industries
(sterilized)
K (cm
-1
) 1.0 0.01
Accuracy < 2% < 2%
Transmitter measurement range 1 μS to 20 mS.cm
-1
0.01 to 200 μS.cm
-1
Pt100 temperature response (t
90%)
< 3 minutes < 45 seconds
Maximum temperature (°C) 125 (8312)
150 (8317)
150
Maximum pressure (bars) 10 (8312)
25 (8317)
25
Sample connection ¾ inch NPT Tri-Clamp 1½ or 2 inch
Table 2 Flow chamber specifications
08313=A=0001 08318=A=0001
Material PVC 316 L SS
Maximum temperature (°C) 60 at 2 bar 150
Maximum pressure (bars) 15 at 25°C 25
Sensor connection ¾ inch NPT ¾ inch NPT
Sample connection ¾ inch NPT ¼ inch NPT
English 3