EasyManua.ls Logo

Hach POLYMETRON 8312 - User Manual

Hach POLYMETRON 8312
456 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
DOC023.98.80083
Polymetron Conductivity
Sensors
02/2017, Edition 3
User Manual
Bedienungsanleitung
Manuale dell'utente
Manuel d'utilisation
Manual del usuario
Manual do utilizador
Uživatelská příručka
Brugsanvisning
Gebruikershandleiding
Instrukcja obsługi
Bruksanvisning
Käyttäjän käsikirja
Ръководство за потребителя
Használati útmutató
Manual de utilizare
Naudotojo vadovas
Руководство пользователя
Kullanım Kılavuzu
Návod na obsluhu
Navodila za uporabo
Korisnički priručnik
Εγχειρίδιο Λειτουργίας
Kasutusjuhend
 

Table of Contents

Questions and Answers

  • Z
    Zachary JenningsAug 17, 2025
    What to do if my Hach Accessories say 'CAL OVERDUE'?
    • J
      Janice DiazAug 17, 2025
      If your Hach Accessories display a 'CAL OVERDUE' message, you should calibrate the sensor.
  • K
    Kristin FisherAug 20, 2025
    What to do if my Hach Accessories say 'NOT CALIBRATED'?
    • M
      Ms. Traci ThompsonAug 20, 2025
      If the Hach Accessories show 'NOT CALIBRATED', you should calibrate the sensor.
  • J
    Justin McculloughAug 23, 2025
    What does 'REPLACE SENSOR' mean on my Hach Accessories, and what should I do?
    • A
      Angela EdwardsAug 23, 2025
      If the Hach Accessories display 'REPLACE SENSOR' because the sensor has been in operation for more than 365 days, calibrate the sensor with a reference solution and reset the sensor days. If the calibration fails, call technical support.
  • R
    Ryan YoungAug 25, 2025
    What to do if my Hach Accessories displays 'ADC FAILURE'?
    • R
      Renee MarshAug 26, 2025
      If the Hach Accessories show 'ADC FAILURE' because the analog to digital conversion failed, ensure that the sensor module is fully inserted into the controller connector. If the issue persists, replace the sensor module.
  • S
    samuel01Aug 27, 2025
    Why is my Hach Accessories showing 'SENS OUT RANGE'?
    • V
      valenciastacyAug 28, 2025
      The Hach Accessories display 'SENS OUT RANGE' because the sensor signal is outside of the accepted limits for the cell constant that is used. To resolve this, make sure that the sensor is configured for the correct cell constant.
  • C
    cherylcarpenterAug 30, 2025
    What to do if my Hach Accessories show 'MEAS TOO HIGH'?
    • D
      dwayneyoderAug 30, 2025
      If the Hach Accessories show 'MEAS TOO HIGH' because the measured value is > 2,000,000 µS/cm, 1,000,000 ppm or 20,000 ppt, make sure that the sensor is configured for the correct cell constant.
  • M
    Melissa DanielSep 1, 2025
    What to do if my Hach Accessories show 'MEAS TOO LOW'?
    • M
      marissa72Sep 2, 2025
      If the Hach Accessories display 'MEAS TOO LOW' because the measured value is < 0 µS/cm, 0 ppm or 0 ppt, make sure that the sensor is configured for the correct cell constant.
  • J
    Jody LawsonSep 4, 2025
    What does it mean when my Hach Accessories say 'OUTPUTS ON HOLD'?
    • S
      Stephanie WagnerSep 4, 2025
      If the Hach Accessories display 'OUTPUTS ON HOLD', the outputs will become active after the selected time period.
  • L
    Leslie HancockSep 6, 2025
    What does 'ZERO TOO HIGH' mean on my Hach Accessories?
    • K
      kaylalesterSep 6, 2025
      If the Hach Accessories display 'ZERO TOO HIGH' because the zero calibration value is > 500,000 counts, make sure that the sensor is held in air during zero calibration and is not located near radio frequency or electromagnetic interference. Also, ensure that the cable is shielded by metal conduit.
  • V
    Veronica GilbertSep 9, 2025
    Why is my Hach Accessories showing 'TEMP TOO HIGH'?
    • E
      Eddie DuffySep 9, 2025
      The Hach Accessories display 'TEMP TOO HIGH' because the measured temperature is > 130 °C. To resolve this, make sure that the sensor is configured for the correct temperature element.

Summary

Product Specifications

General Product Information

Product Safety Information

Information regarding potential hazards, warnings, and precautions for safe operation.

Sensor Installation

Sensor Mounting Procedures

Instructions and warnings for physically installing the sensor, emphasizing safety.

Sensor Module Connection

Sensor Operation

Sensor Configuration Options

Guide to accessing the Configure menu for sensor identification and data handling options.

Sensor Adjustment & Calibration

Zero Calibration Process

Step-by-step guide to perform zero calibration on the conductivity sensor.

Calibration Methods

Reference Solution Calibration

Detailed steps for calibrating the sensor using a reference solution.

Process Sample Calibration

Process Sample Calibration Steps

Steps for calibrating the sensor using a sample from the actual process.

Maintenance and Reset Procedures

Troubleshooting Guide

Conductivity Sensor Resistance Test

Procedure for testing sensor resistance using an ohmmeter to diagnose calibration failures.

Error and Warning Lists

Sensor Error List

Table of common sensor errors, their descriptions, and resolutions.

Parts and Accessories