EasyManua.ls Logo

Hach Polymetron 9610sc - User Manual

Hach Polymetron 9610sc
382 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
DOC023.98.80249
Polymetron 9610sc/9611sc
07/2016, Edition 3
Maintenance and Troubleshooting
Wartung und Fehlerbehebung
Manutenzione e risoluzione dei problemi
Maintenance et dépannage
Mantenimiento y solución de problemas
Manutenção e resolução de problemas
Údržba a odstraňování problémů
Vedligeholdelse og fejlfinding
Onderhoud en probleemoplossing
Konserwacja i usuwanie usterek
Underhåll och felsökning
Huolto ja vianmääritys
Поддръжка и Oтстраняване на проблеми
Karbantartás és hibaelhárítás
Întreţinerea şi depanarea
Techninė priežiūra ir trikčių šalinimas
Техническое обслуживание и устранение неполадок
Bakım ve Sorun Giderme
Údržba a riešenie problémov
Vzdrževanje in odpravljanje težav
Održavanje i rješavanje problema
Συντήρηση και αντιμετώπιση προβλημάτων
Hooldus ja tõrkeotsing

Table of Contents

Questions and Answers

  • M
    Melissa BaileyAug 22, 2025
    Why is the Hach Measuring Instruments reading high?
    • J
      Jeffrey LeeAug 23, 2025
      High readings on Hach Measuring Instruments can stem from these issues: * A leak in a reagent delivery valve. Perform the diagnostic test for reagent delivery valves and replace any leaking valves. * The quantity of reagent supplied to the sample cell is incorrect. Perform the diagnostic test for reagent delivery, and if incorrect, check for blockages in the tubing or replace the solenoid valve. * The reagent blank value in the REAGENTS/STANDARDS menu does not match the value on the R1 (molybdate reagent) bottle. Correct the reagent blank value in the menu to match the bottle. * There is a blue stain on the sample cell. Replace the sample cell and use the reagent set with the modified R2 (citric acid) reagent.
  • R
    reynoldsdianaSep 3, 2025
    What to do if the reagent pressure is low in Hach Measuring Instruments?
    • M
      Misty JarvisSep 4, 2025
      Low reagent pressure in Hach Measuring Instruments can be due to: * A bottle cap not being on tightly or not creating a good seal. Remove the bottle caps, clean the bottle rims, check the inner surfaces of the caps, and fully tighten the caps, ensuring the fittings are also tight. * A leak or bad seal in one of the reagent bottles or tubes. Complete the diagnostic test for low reagent pressure.
  • J
    Johnny BlackNov 10, 2025
    Why does my Hach Measuring Instruments show a calibration error?
    • E
      Erin WilliamsonNov 10, 2025
      A calibration error in Hach Measuring Instruments can occur for several reasons: * The calibration solution value in the menu might not match the value on the bottle. In this case, change the calibration solution value in the calibration menu to match the bottle. * There may be a leak in a reagent delivery valve. Perform the diagnostic test for reagent delivery valves and replace any leaking valves. * The instrument may be supplying an incorrect quantity of reagent or calibration solution to the sample cell. Perform the diagnostic tests for reagent and calibration solution delivery. Look for blockages in the tubing or replace the applicable solenoid valve if needed. * The stir bar may not be installed correctly or is not moving. Install the stir bar and ensure it moves during meas...
  • A
    Andrea DavisNov 15, 2025
    What to do if Hach Measuring Instruments readings are not stable?
    • D
      Deanna Murray DDSNov 15, 2025
      Unstable readings on Hach Measuring Instruments can be caused by several factors: * A leak in one of the reagent delivery valves. Complete the diagnostic test for reagent delivery valves and replace the valve if a leak is found. * An incorrect quantity of reagent is being supplied to the sample cell. Complete the diagnostic test for reagent delivery. If the reagent delivery is incorrect, check for blockages in the tubing or replace the applicable solenoid valve. * Bubbles in the sample cell or on the stir bar. Rinse the sample cell to remove bubbles; if the readings remain unstable, replace the sample cell. If there are bubbles on the stir bar, replace the stir bar. * A blue stain on the sample cell. Replace the sample cell and use the reagent set with the modified R2 (citric acid...
  • D
    David KlineNov 28, 2025
    Why is the Hach Measuring Instruments reading low or less than zero?
    • C
      Cynthia WilliamsDec 1, 2025
      If your Hach Measuring Instruments reading is low or less than zero, consider the following: * A leak in one of the reagent delivery valves. Perform the diagnostic test for reagent delivery valves and replace the applicable reagent delivery valve if a leak is present. * The stir bar is not installed correctly or is not moving. Install the stir bar and ensure it moves during measurements. * The quantity of reagent supplied to the sample cell is incorrect. Perform the diagnostic test for reagent delivery. If the reagent delivery is incorrect, look for a blockage in the tubing or replace the applicable solenoid valve. * The reagent blank value in the REAGENTS/STANDARDS menu is different from the value on the R1 (molybdate reagent) bottle. Change the reagent blank value in the REAGENT...

Hach Polymetron 9610sc Specifications

General IconGeneral
ParameterSilica, Phosphate, Free Chlorine
Measurement MethodColorimetric
Operating Temperature5 - 45 °C
Power Supply100 - 240 VAC, 50/60 Hz
Resolution (Free Chlorine)0.01 mg/L
Accuracy (Silica)±5% of reading or ±5 µg/L, whichever is greater
Cycle Time5 minutes
Response TimeTypically within cycle time
Sample Temperature5 - 40 °C
Sample Flow Rate0.1 - 0.5 L/min
Outputs4-20 mA

Related product manuals