M
Melissa BaileyAug 22, 2025
Why is the Hach Measuring Instruments reading high?
- JJeffrey LeeAug 23, 2025
High readings on Hach Measuring Instruments can stem from these issues: * A leak in a reagent delivery valve. Perform the diagnostic test for reagent delivery valves and replace any leaking valves. * The quantity of reagent supplied to the sample cell is incorrect. Perform the diagnostic test for reagent delivery, and if incorrect, check for blockages in the tubing or replace the solenoid valve. * The reagent blank value in the REAGENTS/STANDARDS menu does not match the value on the R1 (molybdate reagent) bottle. Correct the reagent blank value in the menu to match the bottle. * There is a blue stain on the sample cell. Replace the sample cell and use the reagent set with the modified R2 (citric acid) reagent.






