Robin MillerSep 12, 2025
What does 'WRONG LINEAR TC' mean on my Hach Accessories?
- EElizabeth JohnsonSep 12, 2025
If the Hach Accessories display 'WRONG LINEAR TC', the value must be between 0 and 4%/°C; 0 to 200 °C.
What does 'WRONG LINEAR TC' mean on my Hach Accessories?
If the Hach Accessories display 'WRONG LINEAR TC', the value must be between 0 and 4%/°C; 0 to 200 °C.
What does 'WRONG TC TABLE' mean on my Hach Accessories?
If the Hach Accessories display 'WRONG TC TABLE', the temperature is above or below the temperature range defined by the table.
Brand | Hach |
---|---|
Model | POLYMETRON 8312 |
Category | Measuring Instruments |
Language | English |
Information regarding potential hazards, warnings, and precautions for safe operation.
Instructions and warnings for physically installing the sensor, emphasizing safety.
Guide to accessing the Configure menu for sensor identification and data handling options.
Step-by-step guide to perform zero calibration on the conductivity sensor.
Detailed steps for calibrating the sensor using a reference solution.
Steps for calibrating the sensor using a sample from the actual process.
Procedure for testing sensor resistance using an ohmmeter to diagnose calibration failures.
Table of common sensor errors, their descriptions, and resolutions.