HAMILTON-H900 Quick Guide
Display overview
*dual limb, single use breathing circuits only
HAMILTON-H900
Alarm silence button
1 Press (if appropriate for alarm)
Silence for 2 min
Power button
1 ON: Long-press > 3 s
Device starts in invasive mode
Briefly press to (de)activate standby
2 OFF: Long-press > 3 s
Auto mode
1 Press for default settings
Invasive mode: 37°C
Noninvasive mode: 31°C
Temperature gradient: + 2°C,
(+ 3°C for infant breathing set)
Open temperature monitoring
window
1 Press to show temperature
a) At Y-piece
b) At chamber exit
c) Difference between a) and b)
2 Press again to close window
(Non)invasive mode
1 Press to switch between invasive
and noninvasive modes
Expiratory tube button
1 Press to start extra heating of
expiratory tube*
Reduce condensate in expiratory tube
(white)
2 Press again to stop extra heating
Tube connection indicators
Green: connection correct
Orange: testing
Red: disconnected or faulty
See the HAMILTON-H900 Instructions for use for detailed information.
Question and Answer IconNeed help?

Do you have a question about the Hamilton H900 and is the answer not in the manual?

Hamilton H900 Specifications

General IconGeneral
BrandHamilton
ModelH900
CategoryMedical Equipment
LanguageEnglish

Summary

Display Overview and Controls

Temperature Monitoring Window

Instructions for accessing and closing the temperature monitoring window.

Auto Mode Functionality

Details on default settings for invasive and noninvasive modes.

Power Button Operation

Procedure for turning the device on and off.

Tube Connection Status

Explanation of indicators for tube connection correctness.

Expiratory Tube Heating

How to activate and deactivate extra heating for the expiratory tube.

Ventilation Mode Switching

Method for switching between invasive and noninvasive modes.

Alarm Silence Function

Procedure for silencing alarms for a limited duration.

Circuit and Display Adjustments

Temperature Gradient Adjustment

How to adjust the temperature difference between chamber and Y-piece.

Water Chamber Temperature

How to adjust the water chamber exit temperature.

System Setup and Initial Settings

Connecting System Components

Steps for connecting power, humidifier base, water chamber, and circuits.

Optional Communication Setup

Instructions for connecting a communication cable.

Initial Mode Selection

Guidance on choosing Auto or Manual mode based on conditions.

Optimizing Settings

Recommendations for reducing condensation and setting optimal parameters.

Alarms and Troubleshooting Guide

High-Priority Alarms

Causes and troubleshooting for high-priority alarms like temperature and water level.

Medium-Priority Alarms

Causes and troubleshooting for medium-priority alarms like temperature and water level.

Circuit and Chamber Issues

Troubleshooting for incorrect tube insertion and missing water chamber.

Humidifier Positioning Note

Important note on humidifier positioning to avoid condensation.