Speci cations
40
pH, mV in pH, mV in ORP Temperature
Range
-2.00 to 16.00 pH
-2.000 to 16.000 pH*
±1000.0 mV in pH
±2000.0 mV in ORP
-20.0 to 120.0 ºC ; -4.0 to 248.0 °F**
Resolution
0.01 pH
0.001 pH*
0.1 mV in pH
0.1 mV in ORP
0.1 ºC; 0.1 ºF
Accuracy
@ 25 °C / 77 °F
±0.01pH
±0.002 pH*
±0.2 mV in pH
±0.2 mV in ORP (±999.9 mV)
±1 mV in ORP (±2000 mV)
±0.5 ºC; ±0.9 ºF
pH Calibration
Automatic, up to 3 points (5 points*) calibration, 5 standard (7
standard*) buffers available (1.68*, 4.01 or 3.00, 6.86, 7.01, 9.18, 10.01,
12.45*) and 2 custom buffers*
Relative mV
Calibration
Single point calibration
Temperature
compensation
Automatic -5 to 100º C (23 to 212º F) (using integral temperature sensor)
Log feature
Up to 1000* records organized in:
Log on demand (Max. 200 logs)
Log on stability (Max. 200 logs)
Interval logging*
Additional Speci cations
PC Interface
Micro USB
Storage Interface
USB
Power Supply
5 VDC Adapter (included)
Environment
0-50 ºC (32-122 ºF) Max 95% RH non-condensing
Dimensions
202 x 140 x 12 mm (7.9 x 5.5 z 0.5”)
Weight
250g (8.82 oz)
pH Electrode
HI 11310 Intelligent pH/temperature electrode (included)
Speci cations
*Standard Mode Only
** pH and temperature will be reduced to actual probe/sensor limits.