Press CAL. The instrument will display the measured pH, the LCD first
expected buffer and the temperature reading.
If necessary, press the ARROW keys to select a different buffer value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the selected buffer, Confirm
function key is displayed.
Press Confirm to confirm first point.
The calibrated value and the second expected buffer value is then dis-
played on the LCD.
After the first calibration point is confirmed, immerse the pH electrode and
the temperature probe approximately 3 cm (1¾”) into the second buffer
solution and stir gently. The temperature probe should be close to the pH
electrode.
If necessary, press the ARROW keys to select a different buffer value.
The “ ” tag will blink on the LCD until the reading is stable.