EasyManua.ls Logo

Hanna Instruments HI 2212 - User Manual

Hanna Instruments HI 2212
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
1
www.hannainst.com
pH/mV/°C
Bench Meters
with Custom Buffers
HI 2212 HI 2213
Instruction Manual

Questions and Answers

  • A
    Alicia SchneiderAug 14, 2025
    What to do if Hanna Instruments Measuring Instruments fail to calibrate or give faulty readings?
    • D
      dstevensAug 14, 2025
      If your Hanna Instruments Measuring Instruments fail to calibrate or give faulty readings, the issue might be a broken pH electrode. The suggested solution is to replace the electrode.
  • J
    John RodriguezAug 17, 2025
    What to do if Hanna Instruments HI 2212 display shows pH and -2.00 or 16.00 blinking?
    • S
      scarterAug 17, 2025
      If the display of your Hanna Instruments Measuring Instruments shows "p H" and "-2.00" or "16.00" blinking, it indicates that the reading is out of range in the pH scale. Here are steps to address this: * Verify that the electrode is properly connected. * Ensure that the shipping cap has been removed and recalibrate the meter. Confirm that the pH sample is within the specified range. * Check the electrolyte level and the general condition of the electrode.
  • N
    Natasha ValenzuelaAug 21, 2025
    Why Hanna Instruments HI 2212 readings fluctuate up and down (noise)?
    • K
      Karen FrancisAug 21, 2025
      Fluctuating readings (noise) in Hanna Instruments Measuring Instruments can be due to a clogged or dirty junction, or a low electrolyte level (for refillable electrodes only). To resolve this: * Clean the electrode. * Refill with fresh solution (for refillable electrodes only). * Check the cables and connectors.
  • A
    Amy HillAug 25, 2025
    What to do if Hanna Instruments HI 2212 Measuring Instruments display shows mV and -2000 or 2000 blinking?
    • J
      Jeremy NixonAug 25, 2025
      If the display of your Hanna Instruments Measuring Instruments shows "m V" and "-2000" or "2000" blinking, it means the reading is out of range in the mV scale. Verify that the electrode is connected.
  • D
    Donald GoodmanAug 27, 2025
    What to do if Hanna Instruments HI 2212 Measuring Instruments do not accept the buffer solution for calibration?
    • D
      djohnsonAug 27, 2025
      If your Hanna Instruments Measuring Instruments do not accept the buffer solution for calibration, it could be due to a dirty electrode or contaminated buffer. You should follow the cleaning procedure. If it still fails, replace the electrode and replace the buffer.
  • M
    morganmichaelAug 30, 2025
    How to fix slow response or excessive drift in Hanna Instruments HI 2212 Measuring Instruments?
    • E
      Evan HamiltonAug 31, 2025
      Slow response or excessive drift in Hanna Instruments Measuring Instruments is often caused by a dirty electrode. Clean the electrode and then soak the tip in HI7061 or HI8061 for 30 minutes.
  • T
    Taylor GonzalezSep 3, 2025
    What to do if Hanna Instruments HI 2212 do not work with the temperature probe?
    • C
      Charles LewisSep 3, 2025
      If your Hanna Instruments Measuring Instruments do not work with the temperature probe, it could be due to a broken or wrong temperature probe. Replace the temperature probe.
  • D
    Darrell SmithSep 6, 2025
    What to do if Hanna Instruments HI 2212 display all LCD tags permanently at startup?
    • M
      Mia GordonSep 6, 2025
      If your Hanna Instruments Measuring Instruments displays all LCD tags permanently at startup, one of the keys might be stuck. Check the keyboard or contact the vendor.
  • L
    Leslie SharpSep 9, 2025
    What to do if Hanna Instruments HI 2212 Measuring Instruments display "Errxx" error message?
    • L
      lisajohnsonSep 10, 2025
      If your Hanna Instruments Measuring Instruments display an "Errxx" error message, it indicates an internal error. Power off the meter and then power it on again. If the error persists, contact the vendor.

Summary

WARRANTY

Preliminary Examination

Inspect instrument for shipping damage and check supplied accessories.

General Description

Overview of HI 2212/HI 2213 features like pH, mV, temperature measurement, and GLP.

Functional Description HI 2212 and HI 2213

Details front and rear panel controls and connectors of the meters.

HI 2212 and HI 2213 Specifications

Lists technical specifications like range, resolution, accuracy, and power requirements.

Operational Guide

Power Connection

Instructions for connecting the 12 VDC adapter to the instrument's power supply socket.

Electrode and Probe Connections

Guide for connecting pH/ORP electrodes and temperature probes to the instrument.

Instrument Start-Up

Steps to turn on the instrument and what to expect during the initial self-test.

pH Measurements

Procedure for taking pH measurements after calibration, including sample submersion and stabilization.

mV;ORP Measurements (HI 2213)

Steps for performing mV/ORP measurements using the HI 2213, including electrode preparation and reading stabilization.

Relative mV Measurements (HI 2213)

Procedure for relative mV measurements on the HI 2213, explaining the offset and displayed difference.

Temperature Measurements

How to perform temperature measurements using the HI 7662 probe.

Memory Function

Instructions for storing and recalling measurements using the MEM and MR keys.

pH Calibration

Preparation

Guidelines for preparing buffer solutions and beakers for accurate pH calibration.

Procedure

Overview of calibration procedures, including one, two, or three-point options and custom buffers.

Three-Point Calibration

Detailed steps for performing a three-point pH calibration using buffer solutions.

Two-Point Calibration

Instructions for performing a two-point pH calibration after a three-point calibration.

One-Point Calibration

Details on one-point calibration and its setup options ("Pnt" vs "OFFS").

Notes on Calibration Errors

Explains common error messages like "WRONG" and "BUFFER pH" during calibration.

Working with Custom Buffers

How to select and adjust custom buffer values during pH calibration.

pH Buffer Temperature Dependence

Good Laboratory Practice (GLP)

Expired Calibration

How the instrument alerts users about expired calibration intervals.

pH Calibration Data

How to view stored calibration data including time, date, offset, and slope.

pH calibration buffers

Displays the pH calibration buffers in order and explains "OLd" message.

Calibration Expiration status

Shows how the instrument displays calibration expiration status (OFF, days remaining, or expired).

Calibration alarm

Details how the calibration alarm is displayed and what "no CAL" means.

Setup Mode

Expired Calibration Alarm

How to set or disable the calibration alarm duration (1-7 days).

First Custom Buffer

How to set or modify the value for the first custom pH buffer.

Second Custom Buffer

How to set or modify the value for the second custom pH buffer.

One-Point Calibration Behavior

How to configure the behavior for one-point calibration (Pnt or Offs).

Current Time

How to set the current hour and minute on the instrument.

Current Date

How to set the current year, month, and day on the instrument.

Beep Status

How to enable or disable the audible beep function.

Temperature Unit

How to select the desired temperature unit (°C or °F).

Temperature Calibration

mV Calibration

Relative mV Calibration (HI 2213)

Electrode Conditioning & Maintenance

Preparation Procedure

Steps to prepare electrodes, including removing caps, handling salt deposits, and soaking.

For Refillable Electrodes

Instructions for refilling electrolyte in electrodes and optimizing response time.

For AMPHEL Electrodes

Troubleshooting for AMPHEL® electrodes, suggesting replacement if unresponsive.

Measurement

Proper technique for submerging and stirring electrodes during measurement.

Storage Procedure

Guidelines for storing electrodes to maintain moisture and prevent clogging.

Cleaning Procedure

Detailed procedures for cleaning electrodes for general, protein, inorganic, and oil/grease contamination.

Troubleshooting Guide

Temperature Correlation for pH Sensitive Glass

Accessories

pH Buffer Solutions

Lists available pH buffer solutions in sachets and bottles.

Electrode Storage Solutions

Lists available electrode storage solutions.

Electrode Cleaning Solutions

Lists available electrode cleaning solutions for various contaminants.

Electrode Refill Electrolyte Solutions

Lists available electrolyte solutions for refilling electrodes.

ORP Pretreatment Solutions

pH Electrodes

ORP Electrodes

Extension Cable for Screw-Type Electrodes

Other Accessories

Recommendations for Users

Hanna Instruments HI 2212 Specifications

General IconGeneral
pH Range-2.00 to 16.00 pH
pH Resolution0.01 pH
pH Accuracy±0.01 pH
mV Accuracy±0.2 mV
Temperature Resolution0.1 °C
Temperature Accuracy±0.5 °C
Calibration Pointsup to 5 points
DisplayLCD
ConnectivityUSB
TypepH Meter
mV Resolution0.1 mV
Calibration BufferspH 4.01, 7.01, 10.01
Temperature CompensationAutomatic
Power SupplyAC adapter or batteries
Environment0 to 50 °C
mV Range±2000 mV

Related product manuals