18
SECOND CUSTOM BUFFER
Press CAL when “cb2” is displayed. The custom buffer (disabled –
“no” or 0 to 16 pH) and “CFM will start blinking.
Press the ARROW keys to change the custom buffer value.
Press CFM to save the modified custom buffer value.
Press CAL to escape without saving.
ONE-POINT CALIBRATION BEHAVIOR
Press CAL when “1 Pnt” message is displayed on the secondary LCD.
One of the two options (“Pnt” or “OFFS”) and “CFM will start
blinking (see pH CALIBRATION PROCEDURE for details, page 8).
Press the ARROW keys to toggle between “Pnt” and “OFFS” options.
Press CFM to save the behavior for one-point calibration.
Press CAL to escape without saving.
CURRENT TIME
Press CAL when the current time is displayed. The hour and “CFM will
start blinking.
Press the ARROW
keys to change the hour.
Press RANGE (HI 2213) or u (HI 2212). The minutes and “CFM will
start blinking.
Press the ARROW keys to change the minutes.
Press CFM to save the modified value.
Press CAL to escape without saving.