pH/mV/°C
Bench Meters
with Custom Buffers
HI 2212 HI 2213
Instruction Manual

Questions and Answers

  • S
    stephaniebaileyAug 14, 2025
    What to do if Hanna Instruments Measuring Instruments fail to calibrate or give faulty readings?
    • M
      Melissa GarrettAug 14, 2025
      If your Hanna Instruments Measuring Instruments fail to calibrate or give faulty readings, the issue might be a broken pH electrode. The suggested solution is to replace the electrode.
  • C
    Carla KennedyAug 17, 2025
    What to do if Hanna Instruments HI 2213 display shows pH and -2.00 or 16.00 blinking?
    • J
      Jennifer BrooksAug 17, 2025
      If the display of your Hanna Instruments Measuring Instruments shows "p H" and "-2.00" or "16.00" blinking, it indicates that the reading is out of range in the pH scale. Here are steps to address this: * Verify that the electrode is properly connected. * Ensure that the shipping cap has been removed and recalibrate the meter. Confirm that the pH sample is within the specified range. * Check the electrolyte level and the general condition of the electrode.
  • B
    briannarussellAug 20, 2025
    Why Hanna Instruments HI 2213 readings fluctuate up and down (noise)?
    • G
      Gary VegaAug 21, 2025
      Fluctuating readings (noise) in Hanna Instruments Measuring Instruments can be due to a clogged or dirty junction, or a low electrolyte level (for refillable electrodes only). To resolve this: * Clean the electrode. * Refill with fresh solution (for refillable electrodes only). * Check the cables and connectors.
  • L
    Lisa MillerAug 25, 2025
    What to do if Hanna Instruments Measuring Instruments display shows mV and -2000 or 2000 blinking?
    • S
      Susan KingAug 25, 2025
      If the display of your Hanna Instruments Measuring Instruments shows "m V" and "-2000" or "2000" blinking, it means the reading is out of range in the mV scale. Verify that the electrode is connected.
  • M
    Michael SmithAug 27, 2025
    What to do if Hanna Instruments Measuring Instruments do not accept the buffer solution for calibration?
    • C
      Carl FischerAug 27, 2025
      If your Hanna Instruments Measuring Instruments do not accept the buffer solution for calibration, it could be due to a dirty electrode or contaminated buffer. You should follow the cleaning procedure. If it still fails, replace the electrode and replace the buffer.
  • A
    Angela ParkerAug 31, 2025
    How to fix slow response or excessive drift in Hanna Instruments Measuring Instruments?
    • A
      Amy HicksAug 31, 2025
      Slow response or excessive drift in Hanna Instruments Measuring Instruments is often caused by a dirty electrode. Clean the electrode and then soak the tip in HI7061 or HI8061 for 30 minutes.
  • B
    bethmcintyreSep 3, 2025
    What to do if Hanna Instruments HI 2213 Measuring Instruments do not work with the temperature probe?
    • J
      Joseph BarberSep 3, 2025
      If your Hanna Instruments Measuring Instruments do not work with the temperature probe, it could be due to a broken or wrong temperature probe. Replace the temperature probe.
  • P
    Pamela TaylorSep 6, 2025
    What to do if Hanna Instruments HI 2213 Measuring Instruments display all LCD tags permanently at startup?
    • B
      brett11Sep 6, 2025
      If your Hanna Instruments Measuring Instruments displays all LCD tags permanently at startup, one of the keys might be stuck. Check the keyboard or contact the vendor.
  • M
    Meghan GarciaSep 9, 2025
    What to do if Hanna Instruments HI 2213 Measuring Instruments display "Errxx" error message?
    • A
      Anthony PerezSep 10, 2025
      If your Hanna Instruments Measuring Instruments display an "Errxx" error message, it indicates an internal error. Power off the meter and then power it on again. If the error persists, contact the vendor.

Summary

Warranty Information

HI 2212 and HI 2213 Specifications

Operational Guide

Electrode and Probe Connections

Guides on how to connect pH/ORP electrodes and temperature probes to the instrument.

pH Measurements

Details the procedure for taking pH measurements after calibration, including electrode stabilization.

mV;ORP Measurements (HI 2213)

Explains how to perform Oxidation Reduction Potential (ORP) measurements using the HI 2213 model.

Relative mV Measurements (HI 2213)

Describes how to use and perform relative mV measurements on the HI 2213 instrument.

pH Calibration Procedure

Three-Point Calibration

Step-by-step guide for executing a three-point pH calibration with the instrument.

Two-Point Calibration

Instructions for performing a two-point pH calibration, following the three-point procedure.

One-Point Calibration

Details the process for a single-point pH calibration, including setup options.

Good Laboratory Practice (GLP)

Instrument Setup Configuration

Temperature Calibration (Technical Personnel)

mV Calibration (Technical Personnel)

Relative mV Calibration (HI 2213)

Steps for performing relative mV calibration specific to the HI 2213 model.

Electrode Conditioning and Maintenance

Storage Procedure

Best practices for storing electrodes to maintain their performance and prevent drying.

Cleaning Procedure

Detailed methods for cleaning electrodes based on contamination type (general, protein, inorganic, oil/grease).

Troubleshooting Guide

Recommendations for Users

Hanna Instruments HI 2213 Specifications

General IconGeneral
ModelHI 2213
CategoryMeasuring Instruments
TypepH/ORP Meter
pH Range-2.00 to 16.00 pH
pH Resolution0.01 pH
pH Accuracy±0.01 pH
ORP Resolution1 mV
DisplayLCD
ORP Range±2000 mV
Temperature Resolution0.1 °C
Temperature Accuracy±0.5 °C
CalibrationUp to 5 points
pH CalibrationAutomatic (up to 5 points)
Temperature CompensationAutomatic
Input Impedance10^12 ohms
ORP Accuracy±0.2 mV