Instruction Manual
HI 2221 HI 2223
pH/mV/ºC
Bench Meters
with Calibration
Check
www.hannainst.com
www.ponpe.com
sale@ponpe.com

Questions and Answers

  • M
    Mason PriceAug 12, 2025
    What to do if Hanna Instruments HI 2221 Measuring Instruments do not accept the buffer solution for calibration?
    • C
      Christopher RodriguezAug 12, 2025
      If your Hanna Instruments Measuring Instruments do not accept the buffer solution for calibration, it could be due to a dirty electrode or contaminated buffer. Follow the cleaning procedure. If there are still no results, replace the electrode and the buffer.
  • D
    Daniel ThompsonAug 16, 2025
    What to do if Hanna Instruments Measuring Instruments display shows pH and -2.00 or 16.00 blinking?
    • L
      Larry HerringAug 17, 2025
      If the display of your Hanna Instruments Measuring Instruments shows "pH" and "-2.00" or "16.00" blinking, it means the reading is out of range in the pH scale. To resolve this: * Verify that the electrode is connected. * Verify that the shipping cap has been removed. * Recalibrate the meter. * Make sure the pH sample is in the specified range. * Check electrolyte level and general state of the electrode.
  • L
    Lori WhiteAug 19, 2025
    Why do readings fluctuate up and down on Hanna Instruments HI 2221?
    • A
      Ashley PotterAug 19, 2025
      If the readings on your Hanna Instruments Measuring Instruments fluctuate up and down (noise), it could be due to a clogged or dirty junction, or a low electrolyte level (for refillable electrodes only). Clean the electrode. Refill with fresh solution (for refillable electrodes only). Also, check the cable and connectors.
  • C
    Carol West MDAug 25, 2025
    How to fix slow response or excessive drift in Hanna Instruments HI 2221?
    • D
      Dustin YoungAug 25, 2025
      If you are experiencing a slow response or excessive drift with your Hanna Instruments Measuring Instruments, it is likely due to a dirty electrode. Clean the electrode and then soak the tip in HI7061 or HI8061 for 30 minutes.
  • T
    Todd KingAug 27, 2025
    What to do if Hanna Instruments Measuring Instruments do not work with the temperature probe?
    • G
      griffithwilliamAug 27, 2025
      If your Hanna Instruments Measuring Instruments do not work with the temperature probe, it may be due to a broken temperature probe or using the wrong temperature probe. Replace the temperature probe.
  • S
    Sandra NicholsAug 31, 2025
    What does it mean if Hanna Instruments HI 2221 Measuring Instruments display shows mV and -2000 or 2000 blinking?
    • M
      Melinda MyersAug 31, 2025
      If the display of your Hanna Instruments Measuring Instruments shows "mV" and "-2000" or "2000" blinking, it indicates that the reading is out of range in the mV scale. Ensure that the electrode is properly connected.
  • P
    peter99Sep 4, 2025
    What does "Errxx" mean on Hanna Instruments HI 2221 Measuring Instruments?
    • C
      cervantesnicoleSep 4, 2025
      If your Hanna Instruments Measuring Instruments display an "Errxx" error message, it indicates an internal error. Power off the meter and then power it on. If the error persists, contact the vendor.
  • D
    Daniel LucasSep 7, 2025
    Why does Hanna Instruments HI 2221 display all LCD tags permanently at startup?
    • E
      Eric LopezSep 8, 2025
      If your Hanna Instruments Measuring Instruments display all LCD tags permanently at startup, it indicates that one of the keys is stuck. Check the keyboard or contact the vendor.

Hanna Instruments HI 2221 Specifications

General IconGeneral
ModelHI 2221
CategoryMeasuring Instruments
TypepH/ORP/Temperature Meter
pH Range-2.00 to 16.00 pH
pH Resolution0.01 pH
pH Accuracy±0.01 pH
ORP Range±2000 mV
ORP Resolution1 mV
ORP Accuracy±1 mV
Temperature Resolution0.1 °C
Temperature Accuracy±0.5 °C
Power Supply12 VDC adapter
DisplayLCD
mV Range±2000 mV
mV Resolution1 mV
mV Accuracy±1 mV
CalibrationUp to 5 point calibration
pH CalibrationUp to five-point calibration with seven standard buffers available (pH 1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45)
Temperature CompensationAutomatic or manual
LoggingYes
ConnectivityUSB
Input Impedance10¹² Ohms
Environment0 to 50°C, 95% RH