Immerse the temperature probe or the pH probe including temperature
sensor into the vessel with ice and water as close as possible to the reference
thermometer. Allow a few seconds for the probe to stabilize.
Use the ARROW keys to set the calibration point value to that of ice and
water mixture, measured by the reference thermometer. When the read-
ing is stable and close to the selected calibration point, the Confirm
function key is displayed.
Press Confirm to confirm.
The second expected calibrated point is displayed.
Immerse the temperature probe into the second vessel as close as possible to
the reference thermometer. Allow a few seconds for the probe to stabilize.
Use the ARROW keys to set the calibration point value to that of the hot
water.
When the reading is stable and close to the selected calibration point,
Confirm function key is displayed.
Press Confirm to confirm. The instrument returns to measurement mode.
Note: Use ARROW keys to change calibration point if necessary (±10.0 ºC)
around the point.
If the reading is not close to the selected calibration point, “Wrong”
message will blink. Change the temperature probe and restart calibra-
tion.