EasyManua.ls Logo

Hanna Instruments HI 8510 Series - Ph Values at Various Temperature

Hanna Instruments HI 8510 Series
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
44
pH VALUES AT VARIOUS
TEMPERATURE
Temperature has an effect on pH. The
calibration buffer solutions are effected by
temperature changes to a lesser degree than
normal solutions.
Please refer to the following chart to perform
the pH calibration:
For instance, if the buffer's temperature is
25°C (77°F), calibrate to read 4.01 or 7.01 or
10.01 on the display.
If the buffer's temperature is 20°C, calibrate
to read 4.00/7.03/10.06 on the display.
If the buffer's temperature is 50°C, calibrate
to read 4.06/6.98/9.82 on the display.
TEMP pH VALUES
°C °F 4.01 6.86 7.01 9.18 10.01
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
32
41
50
59
68
77
86
95
104
113
122
131
140
149
158
4.01
4.00
4.00
4.00
4.00
4.01
4.02
4.03
4.04
4.05
4.06
4.07
4.09
4.11
4.12
6.98
6.95
6.92
6.90
6.88
6.86
6.85
6.84
6.84
6.83
6.83
6.84
6.84
6.85
6.85
7.13
7.10
7.07
7.04
7.03
7.01
7.00
6.99
6.98
6.98
6.98
6.98
6.98
6.99
6.99
9.46
9.39
9.33
9.27
9.22
9.18
9.14
9.10
9.07
9.04
9.01
8.99
8.97
8.95
8.93
10.32
10.24
10.18
10.12
10.06
10.01
9.96
9.92
9.88
9.85
9.82
9.79
9.77
9.76
9.75

Related product manuals