16
INTERFERENCES
Interference may be caused by:
Bromine
Iodine
Chlorine Dioxide
Ozone
Oxidized manganese and Chromium
In case of water with hardness greater than 500 mg/L CaCO
3
, shake the sample for approximately
1 minute after adding the reagent.
In case of water with alkalinity greater than 250 mg/L CaCO
3
or acidity greater than 150 mg/L
CaCO
3
, the color of the sample could disappear or develop only partially. To resolve this,
neutralize the sample with diluted HCl or NaOH.
Alternatively, wait for one minute and
just press READ/
/TIMER. In both
cases, the lamp, cuvette and detector
icons will appear on the display, de-
pending on the measurement phase.
The instrument directly displays
concentration in mg/L of free chlorine on
the Liquid Crystal Display.
TOTAL CHLORINE MEASUREMENT
Fill the cuvette up to the mark with 10
mL of unreacted sample and replace the
cap.
Place the cuvette into the holder and en-
sure that the notch on the cap is positioned
securely into the groove.
Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 96104C and is the answer not in the manual?

Hanna Instruments HI 96104C Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 96104C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals