Press RANGE/GLP/ to edit the desired
day (01-31). If the key is kept pressed
the day number is automatically increased.
Note: It is possible to change the editing
from day to year and to month by pressing
READ/ /TIMER.
Press RANGE/GLP/ to edit the desired
month (01-12). If the key is kept pressed
the month number is automatically
increased.
Press RANGE/GLP/ to edit the desired
year (2000-2099). If the key is kept
pressed, the year number is automatically
increased.
When the correct year has been set, press
ZERO/CFM or READ/ /TIMER to
confirm. Now the display will show the
month blinking.
When the correct month has been set,
press ZERO/CFM or READ/
/TIMER
to confirm. Now the display will show
the day blinking.
DATE INPUT
or
or
Then the date of the last calibration (e.g.:
”01.08.2005”) appears on the display, or
“01.01.2005” if the factory calibration was
selected before. In both cases the year
number is blinking, ready for date input.
Note: If the display shows “STD HIGH”, the standard value was too high. If the display shows
“STD LOW”, the standard value was too low. Verify that both CAL CHECK
TM
Standards
HI 95771-11 Cuvets, A and B are free from fingerprints or dirt and that they are
inserted correctly.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 96771C and is the answer not in the manual?

Hanna Instruments HI 96771C Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 96771C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals