METHOD PROCEDURE
• Remove the cuvette.
Add the content of one packet of
HI93701-0 Free Chlorine Reagent.
Replace the plastic stopper and the cap.
Shake gently for 20 seconds.
Insert the cuvette into the holder and
ensure that the notch on the cap is
positioned securely in the groove.
• Press Read. The display will show a 1 minute countdown prior to the measurement.
To skip the timer, press Read twice. When the timer ends, the meter will perform the
reading. The instrument displays the results in mg/L of chlorine (Cl
2
).
INTERFERENCES
Interference may be caused by:
• Bromine, Iodine, Oxidized forms of Chromium and Manganese, Ozone
Hardness greater than 500 mg/L CaCO
3
, to remove the interference shake the
sample for approximately 2 minutes after adding the powder reagent
Alkalinity greater than 250 mg/L CaCO
3
or acidity value greater than 150 mg/L
CaCO
3
, the color of the sample may develop only partially or rapidly fade, to remove
the interference neutralize the sample with diluted HCl or NaOH