Add the content of one packet of
HI 93748A ascorbic acid to each
cuvet, replace the caps and shake
gently.
Add 0.2 mL of the
HI 93748B alkaline-
cyanide reagent
solution to each cu-
vet, replace the
caps and shake gen-
tly.
Add 1 mL of the HI 93748C
0.1% PAN indicator solution
to each cuvet, replace the
caps and shake gently.
Place the cuvet with the reacted deion-
ized water (blank) into the holder
and ensure that the notch on the cap
is positioned securely into the groove.
Wait for 2 minutes and press ZERO. "SIP" will appear during
zeroing.
Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
Insert the second cuvet with the reacted
sample into the instrument.
ZERO
12
3
6
9
2’
HI 93742
Iron LR/Manganese LR