FREE CHLORINE MEASUREMENTFREE CHLORINE MEASUREMENT
FREE CHLORINE MEASUREMENTFREE CHLORINE MEASUREMENT
FREE CHLORINE MEASUREMENT
H
I 9
3
7
0
1
Fill the cuvet up to the mark with
10 mL of unreacted sample and
replace the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and "SIP" will ap-
pear on the display.
Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
Remove the cuvet.
Powder reagents procedure
Add the content of one packet of HI 93701 DPD
reagent. Replace the cap and shake gently.
Reinsert the cuvet into the instrument.
Press READ and the display will
show "SIP" during measure-
ment.
The instrument directly displays concentration in mg/L of free
chlorine on the Liquid Crystal Display.
Liquid reagents procedure
Add 3 drops of HI 93701A-F DPD1 indicator and 3 drops of
HI 93701B-F DPD1 buffer to another cuvet. Shake gently before
adding 10 mL of unreacted sample. Replace the cap and shake
gently again.
ZERO
Free Chlorine
10 mL
x 3 x 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments MULTIPARAMETER ION SPECIFIC METERS and is the answer not in the manual?

Hanna Instruments MULTIPARAMETER ION SPECIFIC METERS Specifications

General IconGeneral
BrandHanna Instruments
ModelMULTIPARAMETER ION SPECIFIC METERS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals