36
TOTAL CHLORINE MEASUREMENTTOTAL CHLORINE MEASUREMENT
TOTAL CHLORINE MEASUREMENTTOTAL CHLORINE MEASUREMENT
TOTAL CHLORINE MEASUREMENT
H
I 9
3
7
1
1
Fill the cuvet up to the mark with 10 mL of
unreacted sample and replace the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and "SIP" will appear on the display.
Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
Powder reagents procedure
Remove the cuvet and add one packet of DPD Total
Chlorine reagent. Replace the cap and shake gently.
Reinsert the cuvet into the instrument.
Wait for 2 minutes and 30 seconds and then press READ. The
display will show "SIP" during measurement.
The instrument directly displays concentration in mg/L of total
chlorine on the Liquid Crystal Display.
ZERO
Total Chlorine
10 mL

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments MULTIPARAMETER ION SPECIFIC METERS and is the answer not in the manual?

Hanna Instruments MULTIPARAMETER ION SPECIFIC METERS Specifications

General IconGeneral
BrandHanna Instruments
ModelMULTIPARAMETER ION SPECIFIC METERS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals