INTERFERENCES
Interference may be caused by:
Bromine
Chlorine
Ozone
Oxidized forms of chromium and manganese
Alkalinity above 300 mg/L and acidity above 150 mg/L as CaCO
3
require neutralization since the color could instantly fade. To resolve
this, neutralize the sample with diluted HCl or NaOH.
In case of water with hardness greater than 500 mg/L CaCO
3
, shake
the sample for approximately 1 minute after adding the reagent.
Iodine