SCPI Programming Reference
HDG2000 Series Arbitrary Waveform Generator 72
This setting is non-volatile; it will not be changed by power cycling or *RST.
CALibration:STORe
Takes calibration constants in volatile memory (CALibration?), and places them in nonvolatile
memory, where they will not be changed by power cycle.
Parameter
Typical Return
none
Store calibration constants into non-volatile memory: CAL:STOR
CALibration:VALue <value>
Specifies the value of the known calibration signal.
Parameter
Typical Return
Numeric, default 0
none
-
This setting is non-volatile; it will not be changed by power cycling or *RST.
5.2.6 COUNter Subsystem
COUNter ON|OFF
COUNter:MEASure?
COUNter:GATEtime
COUNter:GATEtime?
COUNter:TYPE
COUNter ON|OFF
Enable or disable frequency meter.
Parameter
Typical Return
ON|OFF, default OFF
ON , or OFF
COUNter ON
COUNter:MEASure?
Get current counter measurement value.
Parameter
Typical Return
Current frequency or counter
COUNter:GATEtime AUTO|USER1|USER2|USER3|USER4|USER5|USER6
COUNter:GATEtime?
Set GATEtime