33
Resistance
This section describes how to configure 2 and 4-wire resistance measurements from
the front panel.
Step 1: Configure the test leads, as shown below.
2 wire resistance:
4 wire resistance:
Step 2: press the front panel [
Ω
2 w] or [
Ω
4 w]. The following menu appears.
(Ω 4Wmenu does not include Auto Zero)
Step 3: For the HDM3000, use the Up/Down arrow keys to specify the integration time
in the Power Line Cycle Number (PLC) for measurement with Aperture NPLC=10PLC
by default.1, 10 and 100 PLCs provide normal mode (line frequency noise)

Other manuals for Hantek HDM3000