EasyManuals Logo
Home>Hantek>Multimeter>HDM3000

Hantek HDM3000 User Manual

Hantek HDM3000
93 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #71 background imageLoading...
Page #71 background image
68
Precautions for DC measurement
Thermoelectric EMF error
Thermoelectric voltage is the most common error source in low level DC voltage
measurement. Thermal voltages are generated by connecting circuits using
heterogeneous metals at different temperatures. Each connection between the metals will
form a thermocouple which will generate a voltage proportional to the connection
temperature as shown in the table below. You should minimize thermocouple voltage and
temperature variations during low level voltage measurements. The best connection can
be made by pressing the copper wire to the copper wire.
Load error (DC voltage)
Measurement load errors occur when the resistance of the device-under-test (DUT) is a

Other manuals for Hantek HDM3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hantek HDM3000 and is the answer not in the manual?

Hantek HDM3000 Specifications

General IconGeneral
BrandHantek
ModelHDM3000
CategoryMultimeter
LanguageEnglish

Related product manuals