The continuous measurement method is as follows:
≤ 10 Ω
Display measured resistance and buzzer (if
buzzer is enabled.)
10 Ω to 1.2 k
Display measured resistance, no buzzing
> 1.2 k Ω
Display OPEN, no buzzing
The diode
This section describes how to configure diode tests on the front panel.
Step 1: Configure the test leads, as shown below.
Step 2: Press the one on the front panel to open a menu which specifies
whether the DMM will ring to indicate a successful diode test.
The diode measurement method is as follows:
0 to 4.9 V
The voltage is displayed on the front panel and the instrument beeps when the
signal is converted to a threshold of 0.3 to 0.8V. (if beeping is enabled.)
> 5 V
The front panel shows Open
Frequency and period