EasyManuals Logo

Hantek HDM3000 User Manual

Hantek HDM3000
93 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #81 background imageLoading...
Page #81 background image
78
Note: It is not possible to calibrate the diode or continuity separately because these
functions are based on calibrated resistance measurements. In addition, there is no way to
calibrate the gate timing because this function is controlled by digital logic. Capacitance
gain cannot be calibrated.
Precautions for Testing
During self-testing, an error may be raised if an AC signal appears on the input lead. If the
test lead is too long, it can also cause AC signal display.
To optimize performance:
Ensure that the calibration ambient temperature (TCAL) is constant and between 18
and 28 . The ideal calibration temperature should be 23 , fluctuating 2 .
Ensure that the ambient relative humidity is below 80%.
The connecting copper cable requires a preheating time of 90 minutes.
Connect the input cable shields to the ground. Connect the calibrator LO source to the
ground in addition to the where noted in the procedures. It is important to avoid a
ground loop by connecting the LO to the ground at only one point in the circuit.
Because the instrument is capable of very precise evaluation, you must be extra careful to
ensure that the calibration standards and test procedures used do not cause other errors.
Ideally, the criteria used to validate and adjust equipment should be more accurate than
the error specifications for equipment of all sizes.
For 2-wire resistance measurements, take a null measurement with a short-circuited
lead or high precision 4 terminal low thermal short-circuit
to remove wire resistance.
For zero offset calibration, a 4 terminal low thermal short circuit is required.
Recommended testing equipment
The recommended testing equipment for performance verification and calibration is listed
below. If the required equipment is not available, replace it with a calibration standard with
the same accuracy.
application
Recommended equipment
Zero
calibration
High precision 4 terminal low thermal short-
circuit
The DC
voltage
Fluke5720A
The DC
current
Fluke5720A
resistance
Fluke5720A

Other manuals for Hantek HDM3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hantek HDM3000 and is the answer not in the manual?

Hantek HDM3000 Specifications

General IconGeneral
BrandHantek
ModelHDM3000
CategoryMultimeter
LanguageEnglish

Related product manuals