Harobot
[Considerations]
Confirm that the screen is in body temperature mode.
Aim the thermometer probe at the middle of the forehead and keep it vertical, with a distance of
about 3cm ~ 5cm. Press the measurement key, and after about 1 second, the "tick" will display the
measured value, and the test is completed. If the measured value exceeds the temperature alarm
point (the default temperature is 38 ℃), the "tick" will be issued Five beeps for warning.
When the temperature of forehead is greatly affected by the change of ambient temperature,
please aim at the earlobe and measure it.
Tips:
1. Please keep the sensor and probe cavity clean before and after use.
2. Please keep the thermometer in a stable temperature environment. When the ambient
temperature changes greatly, please leave it for about 10 minutes and then measure it.
3. Please do not measure the body temperature immediately after measuring the temperature of
extremely high or very low temperature objects. Please keep it for ten minutes and then measure
it.
4. When the measured object comes from a place with a large difference from the measured
ambient temperature, it shall stay in the test environment for at least five minutes.
5. Try not to measure it under the condition of blowing, drenching, sweating and applying cosmetics
on the forehead. Do not take measurements within 30 minutes after exercise, bathing, or eating.
[Product system settings]
This product provides five function settings, such as temperature unit, tone switch, temperature
alarm point, temperature offset and measurement mode. The measurement mode setting is set
by the mode switch key, and other settings are set in the setting menu.
The comparison table of setting menu is as follows: