Harobot
[Exception message]
Fault description
Resolvent
Screen display
“L0 or“Hi”
1. Check the measurement object. Measurement cannot be guaranteed
under the conditions of forehead blowing, dripping, sweating, applying
cosmetics, etc.
2. Check the temperature offset setting, factory set to 0.0.
3. Check the operating environment. Environmental change has great
influence on measurement.
4. If the ambient temperature changes too much, or the ultra-high
temperature object is directly transferred to the low temperature object,
the measurement difference will occur. It should be used in a relatively
stable environment after 10 minutes of storage to obtain a new test heat
balance.
5. Check the measurement distance (3-5cm).
no response
Reload the battery; check if it operates under the setup menu
No display or abnormal display
Reload battery
Silent tone
Check if the Beep setting is off
Shut down immediately after
power on
Check the battery capacity and reload the battery
[Product performance]
Product model
HRB-E0110Q
Anthropometric range
32.0 ~ 42.5
Measurement accuracy
Within 33.0 ~ 37.0℃,Not more than ±0.2
Display resolution
0.1
Memory
32 group
Type of protection against
electric shock
Internal power equipment
Degree of protection against
electric shock
Type B application part
Protection degree of liquid
inlet
IPX0
Operation mode
Continuous operation
AP / APG device or not
not
Electromagnetic
compatibility-[GB4824-2013]
Group I B class
Permanent equipment or not
not
working condition
Temperature 1-50
relative humidity 85%
battery
DC 3V
2 section AAA battery