DOUBLE LOOP CONTROL
13 of 17
SETTING THE LOW POWER
LIMIT
For routine operation, the low power limit should be set to
0%, as follows:
• Simultaneously press both the Up-arrow S and
the Down-arrow T keys for approximately 3
seconds until the Input Menu prompt InPt appears
in the upper display and the Setup Menu prompt
SEt appears in the lower display:
InPt / Set
• Press the Up-arrow S key twice. 9LbL appears
in the upper display:
9LbL/ SEt
• Press the MODE key nine times. LoP will appear
in the lower display.
• Use the Down-arrow T key to set the low power
limit, shown in the upper display, to 0.
• Press the DSPY key to return to the main display
format.
SETTING THE HIGH
POWER LIMIT
For routine operation, the high power limit should be set to
an initial value of 75% for heating to temperatures above
200ºC or to 35% for temperatures up to 200ºC.
• Simultaneously press both the Up-arrow S and
the Down-arrow T keys for approximately 3
seconds until the Input Menu prompt InPt appears
in the upper display and the Setup Menu prompt
SEt appears in the lower display:
InPt / Set
• Press the Up-arrow S key twice. 9LbL appears
in the upper display:
9LbL/ SEt
• Press the MODE key ten times. HiP will appear in
the lower display.
• Use the Down-arrow T key to set the high power
limit, shown in the upper display, to 35 or 75.
• Press the DSPY key to return to the main display
format.
Note
The high power limit controls the percentage of full power
that is sent out to the heaters. The optimum setting
depends on the set point temperature and on the heated
device. If this high power limit is set too high and the set
point is close to the high temperature limit of the heated
device, the heater(s) may heat too quickly and burn out.
If the high power limit is set too low, the heated device
may heat too slowl
and may not reach the set point.
Adjust this setting as needed from the recommended
initial settings to achieve the desired heating profile.