CT Series /
Alarms
Set Internal Sensor Alarm
To enter main menu:
1. Press & hold "H" + "T" for 3 seconds
2. Choose parameter: Press "T" or "H"
3. Edit parameter: Press "U" or "D"
4. Save changes: Press & hold "T" or "H" for 3 seconds
Symbol
Start Value
Delay Time
Notification
Voice Alarm
hhi
10
20 min.
Alr 5
Continuous
hlo
-10
20 min.
Alr 6
Intermittently
(f)thi
0,8
2 min.
Alr1/3
Intermittently
(f)tlo
-2
30 min.
Alr2/4
Intermittently
Set External Room Sensor Alarm
To adjust:
1. Choose parameter: uS2
2. Select "th" option
If Alr15 appears, USB Connection error
Symbol
Start Value
Delay Time
Notification
Voice Alarm
Low Temperature
(f)rtl
Below 65°F (18°C)
no
Alr10
Intermittently
High Temperature
(f)rth
Over 83°F (28°C)
no
Alr9
Intermittently
Low Humidity
rhl
below 10RH
no
Alr14
Intermittently
High Humidity
rhh
over 85 RH
no
Alr13
Intermittently
Alarm Activation -
Temperature
rAt
on / off
Alarm Activation -
Humidity
rAh
on / off
Page 14 of 23