PROBES AND ACCESSORIES continued
With the DVM-80UVM, if the cable is completely shorted the meter will read the line
voltage times 0.71. For a nominal 7.2kV circuit the meter will read approximately 5.1kV.
Listed here are the actual cable leakage currents for voltage readings on the
DVM-80UVM digital meter: Meter reading of 0.5kV = 8uA
Meter reading of 1.0kV = 16uA
Meter reading of 2.0kV = 31uA
REPAIRS
All repairs and calibration are performed at HD Electric Company.
If any damage is
found, please contact HD Electric Company at
847-473-4980 or saleshde@emerson.com to arrange for service.