MED 1114 2017-12-21MED 1114 2017-12-21 33/40
*n/a: „Not applicable“ in the internally powered mode
Electromagnetic disturbances – Requirements and tests
The device is intended for use in the electromagnetic environment specied
below. The custo
mer or the user of the device should assure that it is used in such
environments.
Statement for the operational
environments
Inside hospitals except for: near active HF
surgical equipment
and the RF shielded room
of an ME system for magnetic resonance
imaging, where the intensity of EM distur-
bances are high.
Performance features of the
ME system that have been
determined to be essential to the
performance
None
Warning
Use of this equipment adjacent to or stacked
with other equipment should be avoided
because it could result in improper operation.
If such use is necessary, this equipment and
the other equipment should be observed to
verify that they are operating normally.
Use of accessories, transducers and cables
other than those specied or provided by the
manufacturer of this equipment could result
in increased electromagnetic emissions or
decreased electromagnetic immunity of this
equipment and result in improper operation.
Portable RF communications equipment
(including peripherals such as antenna cables
and external antennas) should be used no
closer than 30 cm (12 inches) to any part of
the retinoscope, including cables specied
by the manufacturer. Otherwise, degradation
of the performance of this equipment could
result.
A list of all cables, transducers and
other accessories that are relevant
for the EMC compliance
See attachment
Test Compliance
RF emissions CISPR11 Group 1
Class B
Harmonic Emissions Passed*
Voltage Fluctuations/Flicker Passed*