11/56
med 1212 2013- 0 9-10 med 1212 2013- 0 9-10
Technical Specifications
Nominal voltage 3V DC
Nominal current typ. 220mA
Protection class Internal power supply
Application part Type BF
Device classification according to
IEC 62471:2006
Exempt
Battery Typ 2 Alkaline (1.5 V), size AA/LR6 or 1 HEINE mini 2Z
(rechargeable with HEINE
®
mini NT)
Magnification approx. 10x (depending on distance)
Focal Range (correction) approx. +/- 4dpt
Contact plates Silica glass, multi-coated
Environmental conditions
for operation
10°C to 35°C
30% to 90% rel. humidity
800hPa to 1060hPa
Environmental conditions
for storage and transport
-20°C to 50°C
10% to 95% rel. humidity
500hPa to 1060hPa
Dimensions (L x B x H)
for head and handle
165 x 52 x 35 mm³
Weight ca. 195g (incl. 2 batteries)
General Notes and Warnings
Check the correct operation of the device before use! Do not use the device if there are visible signs
of damage.
Do not use the device in hyperbaric chambers, in explosive or oxygen loaded environments!
Do not use the device near strong magnetic fields like MRI scanners!
Do not look into bright light sources by means of loupes. Hazard of glare!
The performance and safety of the device can only be guaranteed when fitted with original HEINE
accessories and HEINE spare parts. Otherwise the warranty is terminated.
Don’t modify the device! If the device is modified despite that, take care that use of the device is safe by
performing investigations and tests.
Your HEINE mini 3000
®
LED F.O. Dermatoscope is a precise optical instrument. Please handle the device
with care and clean it at regular intervals.
Store and use the device in dry and dust-free environments only! If you don’t use the device for a longer
period of time, please remove the batteries in advance.
Electromagnetic Compatibility
Medical electric devices are subject to special precautionary measures with regard to electromagnetic
compatibility (EMC). Portable and mobile high frequency communication equipment can affect medical
electric devices.
This device is intended for use by medical professionals in the electromagnetic environment specified
below. The user of this device should assure that it is used in such an environment.
The use of accessories, converters or cables other than the ones specified by HEINE might lead to increased
emission reduced electrical immunity of the medical equipment.
The device may not be stacked directly near or used directly beside other devices. If the device is to be
operated in a stack or with other devices, the device should be watched to ensure it operates properly in
this location.

Other manuals for HEINE mini 3000