TROUBLE SHOOTING CHAPTER 8
8 Trouble Shooting
- Erroneous graphical appearance, e.g. some elements missed.
Close PamWin-3.
Switch off PAM-2500.
Delete (or better rename) <PAM_WIN.INI> which is in the
C:\PamWin_3\Data_2500 directory. Start instrument and
software.
- Unable to start computer to PAM-2500 communication.
Close PamWin-3 AND switch off the PAM-2500 fluorome-
ter.
After several seconds, switch PAM-2500 on again, make sure
that the green signal LED flashes and start software.
103