31
Symptom Explanation Action
Err60 General hardware error. a. Turn off the analyzer and turn it on again after
30 seconds. Take a new microcuvette and repeat the
measurement, as described in the Measuring section.
b. If the problem continues, the analyzer needs service.
Contact your local distributor.
Err61 Self test error during start up of analyzer. a. Turn off the analyzer and turn it on again after
30 seconds. Take a microcuvette and perform the
measurement, as described in the Measuring section.
b. If the problem continues, the analyzer needs service.
Contact your local distributor.
Err62 1. Blanking test failed.
2. Optical parts dirty
1. Turn off the analyzer and turn it on again after
30 seconds. Take a new microcuvette and repeat the
measurement, as described in the Measuring section.
2a. Turn off the analyzer and clean the optical parts as
described in the Maintenance section.
2b. If the problem continues, the analyzer needs
service. Contact your local distributor
WbC Empty cuvette holder. Take a microcuvette and perform the measurement, as
described in the Measuring section.
LLL 1. Measured value is below 0.3 x10
9
/L
(300/mm
3
, 300/µL).
2. Empty microcuvette, not filled with
sample.
1. NOTE! Results exceeding the LLL limit should be
verified with a suitable laboratory method and be
questioned as to the pathological condition of the
patient.
2. Take a new microcuvette and repeat the
measurement, as described in the Measuring section
and make sure that the microcuvette is filled with
sample.
HHH Measured value exceeds 30.0 x10
9
/L
(30000/mm
3
, 30000/µL).
NOTE! Results exceeding the HHH limit should be
verified with a suitable laboratory method and be
questioned as to the pathological condition of the
patient.