page Testing Safety Cabinet Class II according to EN 12 469
Installation test Routine maintenance test
Laboratory Class Complies with installation rules
Safety Declaration filed
Yes : No: Not required:
Visual check
Tested
O
Remarks
Values
Visual check
Tested
O
Remarks
Values
Type label DD
Direct with exhaust
air tra
GS – label TD Draught interrupt
Brief operating instructions KÜHLER Exhaust cooler
Warnings AEF
2
nd
exhaust air filter
Attach test label
Check suction protection,
exchan
Check joints of sealing agents Height of inflow opening
Check surface for ruptures
Function monitor contact
Safety check VDE 0701-1 :2000-09 Safety check VDE 0701-1 :2000-09
Ground conductor
< 0,3 ȍ
Conductance
< 3.5 mA
Insulating resistance
> 2 Mȍ
Monitoring equipment, acoustic and visual
Contactor installed in the instrument
Yes No
Contactor for upper limit responds
Yes No
Contactor for lower limit responds
Yes No
Front panel position - Alarm triggers if front panel outside working position
Yes No
Remarks:
Test Record
Date :2001- 10-01